Chapter 20 · Experimental Skills

20.1 Measuring Instruments and Mechanics and Heat Experiments

The JEE Main experiments in measurement, mechanics and heat: Vernier callipers and the screw gauge with zero errors, the decaying pendulum, the metre-scale balance, Young's modulus, capillary rise, terminal velocity, the resonance tube and the method of mixtures — what you do, what you record, which graph you draw and which errors matter.

Every unit of the JEE Main syllabus ends in a laboratory. The paper does not ask you to set up apparatus, but it does ask what you would see on a Vernier scale with a zero error, which graph turns pendulum readings into a straight line, why the marks in a viscosity jar are drawn well below the surface, how two resonance lengths cancel the end correction, or which single measurement controls the error in Young's modulus. These questions are quick marks for anyone who has thought through the experiments — and traps for anyone who has only memorised formulas.

This section covers the nine experiments on measurement, mechanics and heat in the syllabus. The physics behind each one was taught earlier (we link to it rather than repeat it); here each experiment gets the same treatment: aim, apparatus, method, observation table, graph or calculation, and precautions and sources of error, followed by worked JEE-style problems. Reading the instruments comes first, because every other experiment depends on it. Section 1.1 introduced least counts and Section 1.3 taught the Vernier and screw-gauge readings and the idea of zero error; we now use the instruments the way the experiments do — inside and outside diameters, the depth of a vessel, and zero errors of both signs.

Vernier callipers and the screw gauge in the laboratory

Vernier callipers have three measuring parts that move together with the sliding Vernier: the large outside jaws grip an object from outside (external diameter, length), the small inside jaws open against the inner walls of a hollow object (internal diameter), and the thin depth rod slides out of the end of the main scale (depth of a vessel or of a hole). All three share the same Vernier, so they share the same least count and the same zero error. A reading is $\text{MSR}+k\times\text{LC}$, as in Section 1.3; the only new skill is reading the zero error when it is negative.

📐 Rule 20.1 — Zero error of Vernier callipers

Close the jaws fully. For a Vernier with $n$ divisions and least count LC:

  • if the Vernier zero lies to the right of the main-scale zero and the $k$-th Vernier division coincides with a main-scale mark, the callipers read too much: positive zero error $e_0=+k\times\text{LC}$;
  • if the Vernier zero lies to the left of the main-scale zero and the $k$-th division coincides, the callipers read too little: negative zero error $e_0=-(n-k)\times\text{LC}$.
$$\text{corrected reading}=\text{observed reading}-e_0 .$$

Why $-(n-k)$? When the Vernier zero sits a distance $x$ to the left of the main zero, its $k$-th mark sits at $-x+k\,(1-\text{LC})$ MSD. For that mark to land on a main-scale mark (the $(k-1)$-th, just past zero) we need $x=(n-k)\times\text{LC}$ — counting the divisions not yet matched. A common slip is to write $e_0=-k\times\text{LC}$.

Experiment 1 — Internal and external diameter and depth of a vessel

Aim. To measure the external diameter, the internal diameter and the depth of a small beaker (or calorimeter) with Vernier callipers, and hence its wall thickness and internal volume. Apparatus. Vernier callipers, beaker, magnifying glass.

Method.

  1. Find the least count ($1\ \text{MSD}/n$) and read the zero error with the jaws closed (Rule 20.1).
  2. External diameter: grip the beaker between the outside jaws, gently, at right angles to its axis. Read MSR and the coinciding Vernier division. Turn the beaker through $90^\circ$ and repeat (the vessel may not be perfectly round).
  3. Internal diameter: open the inside jaws against the inner wall at the widest position (move the jaws a little to find the maximum reading). Repeat in a perpendicular direction.
  4. Depth: rest the end of the main scale on the rim and slide the depth rod down until it just touches the bottom, holding the scale vertical. Repeat at two or three places.
QuantityMSR (mm)coinciding division $k$observed (mm)corrected (mm)
external diameter$64$$9$$64.45$$64.65$
internal diameter$60$$13$$60.65$$60.85$
depth$72$$4$$72.20$$72.40$

Sample readings from Example 1: $\text{LC}=0.05\ \text{mm}$, $e_0=-0.20\ \text{mm}$ (one reading of each shown; in practice take the mean of several).

Calculation. Wall thickness $=\tfrac12(D_\text{ext}-D_\text{int})$; internal volume (capacity) $V=\tfrac{\pi}{4}D_\text{int}^2\,h$ for a cylindrical vessel.

Precautions and sources of error. Do not press the jaws hard (the vessel squeezes and the jaws tilt); keep the callipers perpendicular to the axis; read with the eye directly above the coinciding marks (parallax); always correct for zero error with its sign. The depth reading is wrong if the main scale is tilted or the rod does not reach the lowest point of a curved bottom.

A screw gauge works the same way with a screw. Bring the flat faces of the anvil and spindle together (using the ratchet) and look at the circular scale against the reference (index) line of the linear scale.

📐 Rule 20.2 — Zero error of a screw gauge

With the faces just touching, for a circular scale of $N$ divisions and least count $\text{LC}=\text{pitch}/N$:

  • if the zero of the circular scale lies below the reference line (it has gone past it) and the $k$-th division is on the line, the gauge reads too much: positive zero error $e_0=+k\times\text{LC}$;
  • if the zero of the circular scale lies above the reference line (it has not reached it) and the $k$-th division is on the line, the gauge reads too little: negative zero error $e_0=-(N-k)\times\text{LC}$.

As for the Vernier, corrected reading $=$ observed reading $-\,e_0$.

📘 Definition 20.1 — Backlash error

Backlash error is the error that arises because a worn screw does not move at once when its direction of rotation is reversed: the head turns through a small angle (and the circular scale changes) before the spindle starts to move back. It is avoided by always turning the screw in the same direction when closing it on the object, and the ratchet slips once the faces touch the object, so the object is never over-tightened.

Experiment 2 — Diameter of a wire and thickness of a sheet with a screw gauge

Aim. To measure the diameter of a given wire and the thickness of a given sheet, and the area of cross-section of the wire. Apparatus. Screw gauge, wire, metal sheet (or a glass slide).

Method.

  1. Find the pitch (distance moved in, say, $5$ rotations $\div5$) and the least count ($\text{pitch}/N$). Find the zero error (Rule 20.2), turning the screw only by the ratchet.
  2. Place the wire between the anvil and spindle and close them by the ratchet until it clicks. Note the linear scale reading (LSR, the last mark visible, including half-millimetre marks) and the circular division on the reference line.
  3. Take readings at four or five places along the wire and, at each place, in two perpendicular directions — this averages out a wire that is not perfectly round or uniform.
  4. For the sheet, take readings at several places spread over its surface.
ObjectLSR (mm)circular divisionobserved (mm)corrected (mm)
wire, position 1$1.0$$27$$1.27$$1.31$
sheet, position 1$2.5$$8$$2.58$$2.62$

Sample readings from Example 2: pitch $0.5\ \text{mm}$, $N=50$, $\text{LC}=0.01\ \text{mm}$, $e_0=-0.04\ \text{mm}$.

Calculation. Mean corrected diameter $d$; area of cross-section $\tfrac{\pi}{4}d^2$, whose percentage error is twice that of $d$ (Section 1.3).

Precautions and sources of error. Use the ratchet (never the thimble) to close the gauge; turn in one direction only (backlash); keep the wire straight and free of kinks; take the mean of many readings; apply the zero error with its sign.

✏️ Example 1 — Measuring a beaker
On a pair of Vernier callipers $20$ Vernier divisions equal $19$ main-scale divisions of $1\ \text{mm}$. With the jaws closed, the Vernier zero lies to the left of the main-scale zero and the $16$th Vernier division coincides with a main-scale mark. A beaker gives these readings — outside jaws: MSR $64\ \text{mm}$, $9$th division coincides; inside jaws: MSR $60\ \text{mm}$, $13$th division; depth rod: MSR $72\ \text{mm}$, $4$th division. Find (a) the least count and the zero error, (b) the corrected external and internal diameters and depth, (c) the wall thickness and (d) the capacity of the beaker in $\text{cm}^3$.
(a) $\text{LC}=\dfrac{1\ \text{mm}}{20}=0.05\ \text{mm}$. Vernier zero to the left, $16$th division coincides: $e_0=-(20-16)\times0.05=-0.20\ \text{mm}$.
(b) Observed: $64+9\times0.05=64.45$, $60+13\times0.05=60.65$, $72+4\times0.05=72.20\ \text{mm}$. Corrected (add $0.20$): $D_\text{ext}=64.65\ \text{mm}$, $D_\text{int}=60.85\ \text{mm}$, $h=72.40\ \text{mm}$.
(c) Wall thickness $=\tfrac12(64.65-60.85)=1.90\ \text{mm}$ — the zero error cancels in a difference.
(d) $V=\tfrac{\pi}{4}D_\text{int}^2h=\tfrac{\pi}{4}(6.085)^2(7.240)=211\ \text{cm}^3$.
Check All three readings use the same Vernier, so the same $e_0$ applies to each. The capacity of about $210\ \text{mL}$ is sensible for a beaker $6\ \text{cm}$ wide and $7\ \text{cm}$ deep.
✏️ Example 2 — A wire and a sheet with a negative zero error
A screw gauge has a pitch of $0.5\ \text{mm}$ and $50$ divisions on its circular scale. With the faces in contact, the zero of the circular scale lies above the reference line and the $46$th division is on the line. For a wire, the linear scale shows $1.0\ \text{mm}$ and the $27$th circular division is on the line; for a sheet, $2.5\ \text{mm}$ and the $8$th division. Find (a) the least count and zero error, (b) the corrected diameter and thickness and (c) the area of cross-section of the wire.
(a) $\text{LC}=0.5/50=0.01\ \text{mm}$. Zero above the line: $e_0=-(50-46)\times0.01=-0.04\ \text{mm}$.
(b) Wire: observed $1.0+27\times0.01=1.27\ \text{mm}$, corrected $1.27+0.04=1.31\ \text{mm}$. Sheet: observed $2.58\ \text{mm}$, corrected $2.62\ \text{mm}$.
(c) $A=\tfrac{\pi}{4}(1.31)^2=1.35\ \text{mm}^2=1.35\times10^{-6}\ \text{m}^2$.
Check Writing $e_0=-0.46\ \text{mm}$ (reading the $46$ directly) would give $1.73\ \text{mm}$ — an error far larger than the gauge's whole purpose. A zero error is always a few least counts.

In Interactive 1 choose Vernier — external diameter, set the zero error to $-0.4\ \text{mm}$ and drag the true diameter to $52.7\ \text{mm}$. What will the measuring panel show?

The zero-check panel shows the Vernier zero to the left of the main zero with division $6$ coinciding: $e_0=-(10-6)\times0.1=-0.4\ \text{mm}$. With a negative zero error the callipers read too little, so the measuring panel shows MSR $52\ \text{mm}$ and division $3$: observed $52.3\ \text{mm}$, and the readouts give corrected $=52.3-(-0.4)=52.7\ \text{mm}$. ($53.1$ would be the result of adding the error to the true value the wrong way; division $6$ is the zero-check coincidence, not the measurement.)
Interactive 1 — Reading Instruments with a Zero Error

Each instrument is shown twice, greatly magnified: panel 1 with the jaws (or studs) closed, to read the zero error, and panel 2 measuring the object. Choose what to measure, set the zero error and the true size, and read both panels before you look at the readouts. The coinciding marks are drawn in red. Press Play to let the size drift.

−0.4 mm52.7 mm
least count—
zero error $e_0$—
observed—
corrected—

A screw gauge has pitch $1\ \text{mm}$ and $100$ circular divisions. With the faces touching, the zero of the circular scale is $3$ divisions above the reference line. A wire reads $2\ \text{mm}$ on the linear scale and $45$ on the circular scale. Its diameter is

The simple pendulum, the metre scale and Young's modulus

A real pendulum does not swing for ever: air drag and friction at the support take a little energy on every swing, and the amplitude shrinks. The syllabus experiment measures that loss. The idea needed is from Section 10.2 (Result 10.10): in SHM the total energy is $E=\tfrac12m\omega^2A^2$, so energy is proportional to the square of the amplitude.

📘 Definition 20.2 — Damped oscillation

An oscillation whose amplitude decreases with time because energy is removed by friction or drag is a damped oscillation. When the damping is light (the amplitude changes very little in one period), the motion is still very nearly SHM with the same period, and its energy at any moment is that of SHM with the current amplitude.

📐 Result 20.3 — Energy of a lightly damped pendulum

For a simple pendulum of length $L$ and bob mass $m$ swinging with small horizontal amplitude $A$ (so $\omega^2=g/L$),

$$E=\tfrac12\,\frac{mg}{L}\,A^2\ \propto\ A^2 .$$

The fraction of the energy remaining at time $t$ is $\left(A/A_0\right)^2$. When the drag is proportional to the speed, equal time intervals remove equal fractions of the energy: $A^2=A_0^2e^{-t/\tau}$, so $A^2$ falls by the same factor in every equal interval of time (and $A$ by the square root of that factor).

Experiment 3 — Dissipation of energy of a simple pendulum

Aim. To study the dissipation of energy of a simple pendulum by plotting a graph between the square of the amplitude and time. Apparatus. Clamp stand, heavy bob on a thread about $1\ \text{m}$ long, a horizontal metre scale fixed just below the bob (or a protractor at the support), stopwatch.

Method.

  1. Set the scale so that its zero (or a reference mark) is exactly below the bob at rest.
  2. Pull the bob aside by a small amplitude $A_0$ (about $10\ \text{cm}$ for a $1\ \text{m}$ pendulum, so that the swing is nearly SHM) and release it without pushing. Start the stopwatch.
  3. Every minute, note the farthest point the bob reaches on the scale — the amplitude $A$ — sighting vertically to avoid parallax.
  4. Continue for $8$–$10$ minutes. Tabulate $t$, $A$ and $A^2$ and plot $A^2$ (proportional to the energy) against $t$.
$t$ (min)$0$$1$$2$$3$$4$$5$
$A$ (cm)$10.0$$9.0$$8.1$$7.3$$6.6$$5.9$
$A^2$ (cm²)$100$$81.0$$65.6$$53.3$$43.6$$34.8$

Graph. $A^2$ against $t$ is a smoothly falling curve, steepest at the start: the energy loss per minute is largest when the energy (and speed) is largest. Its slope at any time, multiplied by $\tfrac12mg/L$, is the rate of energy loss. (Plotting $\ln A^2$ against $t$ would give a straight line of slope $-1/\tau$.)

Precautions and sources of error. Keep the amplitude small; release the bob without a push and without twisting; make sure the support is rigid (a shaking stand takes energy too); read the extreme position quickly and from directly above. A light or large bob (a card attached to it) loses energy much faster because air drag grows with area.

✏️ Example 3 — How much energy has the pendulum lost?
Use the readings in the table of Experiment 3, for a bob of mass $0.20\ \text{kg}$ on a pendulum of length $1.0\ \text{m}$. (a) Show that $A^2$ falls by about the same factor each minute. (b) What fraction of the starting energy is left after $5$ minutes? (c) Find the energy lost in $5$ minutes and the average rate of loss.
(a) $81.0/100=0.81$, $65.6/81.0=0.81$, $53.3/65.6=0.81$, $43.6/53.3=0.82$, $34.8/43.6=0.80$: a constant factor of about $0.81$ per minute — equal fractions of the energy in equal times (and $A$ falls by $\sqrt{0.81}=0.9$ per minute).
(b) $\dfrac{E}{E_0}=\left(\dfrac{5.9}{10.0}\right)^2=0.348$, about $35\%$.
(c) $E_0=\tfrac12\dfrac{mg}{L}A_0^2=\tfrac12\times\dfrac{0.20\times9.8}{1.0}\times(0.100)^2=9.8\times10^{-3}\ \text{J}$. Lost: $9.8\times10^{-3}\times(1-0.348)=6.4\times10^{-3}\ \text{J}$; average rate $\dfrac{6.4\times10^{-3}}{300}=2.1\times10^{-5}\ \text{W}$.
Check After $5$ min the amplitude has fallen only to $59\%$ but the energy to $35\%$ — always square the amplitude ratio.

In Interactive 2 set $A_0=12\ \text{cm}$ and the energy half-life to $4.0\ \text{min}$, then move the clock to $t=8.0\ \text{min}$. What is the amplitude then?

At $t=8.0\ \text{min}$ the readouts show $A=6.00\ \text{cm}$, $A^2=36.0\ \text{cm}^2$ and energy $25.0\%$ of the start. Two half-lives have passed, so $A^2$ (the energy) has halved twice to a quarter, $144\to36\ \text{cm}^2$; the amplitude is the square root, $\sqrt{36}=6\ \text{cm}$ — half the starting amplitude. ($3\ \text{cm}$ treats the amplitude as halving every half-life; $8.5\ \text{cm}$ is one half-life.)
Interactive 2 — Pendulum Amplitude Logger: $A^2$ against Time

A pendulum swings over a centimetre scale; the red arrows mark its current amplitude $A$. Every minute the logger plots a point of $A^2$ (proportional to the energy) against time. Set the starting amplitude and the energy half-life (short for a light bob with a card attached, long for a heavy brass bob), then press Play or drag the clock. The dashed lines mark where the energy has halved.

12.0 cm4.0 min8.0 min
time $t$—
amplitude $A$—
$A^2$—
energy left $E/E_0$—

Experiment 4 — Mass of a body using a metre scale (principle of moments)

Aim. To find the mass of a given body using a metre scale balanced on a knife edge. Apparatus. Metre scale, knife edge (wedge) on a stand, a known mass (slotted weight), the unknown body, two thread loops.

The theory is the principle of moments for a body in equilibrium (Section 5.2, Law 5.8): about the knife edge, the anticlockwise moments balance the clockwise moments — including the moment of the scale's own weight, which acts at its centre of gravity.

📐 Result 20.4 — The metre-scale balance

A metre scale of mass $m_s$ with centre of gravity at $G$ rests on a knife edge at $K$. A known mass $m$ hangs at distance $x_1$ on one side of $K$ and the unknown mass $M$ at distance $x_2$ on the other. In balance,

$$M\,x_2=m\,x_1\quad\text{if }K\text{ is at }G;\qquad M\,x_2=m\,x_1\pm m_s\,|KG|\quad\text{otherwise},$$

with the sign $+$ if $G$ lies on the same side of $K$ as $m$, and $-$ if on the same side as $M$. All distances are measured from the knife edge; the masses cancel $g$.

Method.

  1. Balance the scale alone on the knife edge and note the balance point $G$ (ideally $50.0\ \text{cm}$; often a little off).
  2. Keeping the knife edge at $G$, hang the known mass on one side and the body on the other with thin thread loops. Move one of them until the scale balances horizontally. Record both positions.
  3. Repeat for three or four different positions of the known mass and average the results for $M$.
Known mass $m$ (g)$x_1$ (cm)$x_2$ (cm)$M=mx_1/x_2$ (g)
$50$$30.0$$20.1$$74.6$
$50$$40.0$$26.7$$74.9$

Precautions and sources of error. Use thin, light thread loops (their mass adds to the masses); read each position at the thread, not at the edge of the weight; make sure the scale is truly horizontal and not touching anything; the knife edge must be sharp. If the knife edge is not at $G$ and the scale's weight is ignored, the result is systematically wrong.

✏️ Example 4 — When the knife edge is not at the centre
A uniform metre scale (centre of gravity at $50.0\ \text{cm}$) is balanced on a knife edge at the $40.0\ \text{cm}$ mark when a $50\ \text{g}$ mass hangs at the $20.0\ \text{cm}$ mark. (a) Find the mass of the scale. (b) The $50\ \text{g}$ mass is removed and the knife edge is moved to the $60.0\ \text{cm}$ mark. A body of unknown mass $M$ hung at the $90.0\ \text{cm}$ mark keeps the scale balanced. Find $M$.
Find $M$ in (b) before you open the solution:
(a) Moments about $40.0\ \text{cm}$: the $50\ \text{g}$ mass is $20.0\ \text{cm}$ to the left, the scale's weight $10.0\ \text{cm}$ to the right: $50\times20.0=m_s\times10.0$, so $m_s=100\ \text{g}$.
(b) About $60.0\ \text{cm}$: the scale's weight acts $10.0\ \text{cm}$ to the left, $M$ acts $30.0\ \text{cm}$ to the right. $M\times30.0=100\times10.0$, so $M=33.3\ \text{g}$.
Check The scale itself acts as the "known mass" in (b) — a balance can weigh with no slotted weights at all, once $m_s$ and $G$ are known.
⏱ Shortcut Replace the whole scale by a point mass $m_s$ at $G$, then take moments about the knife edge — $g$ cancels, so work in grams and centimetres.
📘 Definition 20.3 — Searle's apparatus

Searle's apparatus measures the small extension of a long wire. Two identical wires hang side by side from the same rigid support: a reference wire carrying a fixed load and the main scale, and an experimental wire carrying the Vernier (or a micrometer screw with a spirit level) and the variable load. Any sagging of the support or change of temperature affects both wires equally and cancels; only the extra extension of the experimental wire is read.

Experiment 5 — Young's modulus of the material of a wire

Aim. To determine Young's modulus of the material of a given wire using Searle's apparatus. Apparatus. Searle's apparatus, slotted half-kilogram weights, screw gauge, metre scale.

From Section 7.1 (Result 7.2), a wire of length $L$ and radius $r$ stretched by a load $Mg$ extends by $l=\dfrac{MgL}{\pi r^2Y}$.

📐 Result 20.5 — Young's modulus from Searle's experiment
$$Y=\frac{MgL}{\pi r^2\,l}=\frac{gL}{\pi r^2}\times\frac{1}{\text{slope}},\qquad\frac{\Delta Y}{Y}=\frac{\Delta L}{L}+2\frac{\Delta r}{r}+\frac{\Delta l}{l},$$

where the slope is that of the straight-line graph of extension $l$ against load mass $M$ (within the elastic limit). The errors in $r$ (doubled, and $r$ is tiny) and in $l$ (small) dominate; the error in $L$ (about $1\ \text{mm}$ in $2\ \text{m}$) is negligible.

Method.

  1. Hang a dead load (about $1\ \text{kg}$) on both wires to straighten them. Measure $L$ with a metre scale and the diameter at several places with a screw gauge.
  2. Note the Vernier (or micrometer) reading. Add $0.5\ \text{kg}$ at a time, waiting a minute after each addition, and note the reading each time up to about $3$–$4\ \text{kg}$ (well within the elastic limit).
  3. Remove the loads one by one and read again. Use the mean of the loading and unloading readings for each load.
  4. Plot extension against load; find the slope and use Result 20.5.
Load (kg)reading, loading (mm)reading, unloading (mm)mean (mm)extension (mm)
$0$$0.00$$0.01$$0.005$$0$
$1.0$$0.50$$0.51$$0.505$$0.50$
$2.0$$1.00$$1.01$$1.005$$1.00$
$4.0$$2.00$$2.00$$2.000$$2.00$

Precautions and sources of error. Remove kinks with the dead load first; add loads gently and wait for the wire to settle; stay well inside the elastic limit (loading and unloading readings must agree — if they do not, the wire has taken a permanent set); measure the diameter carefully, in perpendicular directions and at many places, because it enters squared.

✏️ Example 5 — Young's modulus and its error
In Searle's experiment a steel wire is $2.00\ \text{m}$ long and its diameter, measured with a screw gauge of least count $0.01\ \text{mm}$, is $0.50\ \text{mm}$. A load of $4.0\ \text{kg}$ produces an extension of $2.00\ \text{mm}$, read with a micrometer of least count $0.01\ \text{mm}$. The length is measured to $0.1\ \text{cm}$; the mass is exact. (a) Find Young's modulus. (b) Find the maximum percentage error in $Y$ and write the result with its uncertainty. (c) Which measurement should the student improve first?
Find the percentage error in (b) first:
(a) $r=0.25\ \text{mm}$, $\pi r^2=1.96\times10^{-7}\ \text{m}^2$. $Y=\dfrac{4.0\times9.8\times2.00}{1.96\times10^{-7}\times2.00\times10^{-3}}=2.0\times10^{11}\ \text{Pa}$.
(b) $\dfrac{\Delta Y}{Y}=\dfrac{0.1}{200}+2\times\dfrac{0.01}{0.50}+\dfrac{0.01}{2.00}=0.05\%+4.0\%+0.5\%=4.55\%$. $\Delta Y=0.0455\times2.0\times10^{11}=0.09\times10^{11}\ \text{Pa}$, so $Y=(2.00\pm0.09)\times10^{11}\ \text{Pa}$.
(c) The diameter contributes $4.0\%$ of the $4.55\%$: measure it with a finer gauge or average many more readings.
⏱ Shortcut In any "which error dominates" question, double the fractional error of each squared quantity and compare: here $2\times\tfrac{1}{50}$ beats $\tfrac{1}{200}$ easily.

In a pendulum experiment the amplitude falls from $10\ \text{cm}$ to $7.0\ \text{cm}$. The fraction of the original energy that has been dissipated is about

Surface tension and viscosity

Both experiments in this subsection measure small lengths in a liquid — a capillary rise of a few centimetres, a bore of a fraction of a millimetre, the fall of a ball between two marks — and both turn on a formula from Chapter 7. The new instrument is the travelling microscope.

📘 Definition 20.4 — Travelling microscope

A travelling microscope is a low-power microscope with cross-wires in its eyepiece, mounted on a carriage that slides along a vertical (or horizontal) main scale with a Vernier. Focusing the cross-wire on two points in turn and taking the difference of the two scale readings gives the distance between them, typically to $0.01\ \text{mm}$ ($0.001\ \text{cm}$). The carriage is always moved in one direction during a set of readings, to avoid backlash.

Experiment 6 — Surface tension of water by capillary rise; effect of detergents

Aim. To determine the surface tension of water by capillary rise and to study the effect of a detergent on it. Apparatus. Clean glass capillary tube, beaker of distilled water, clamp stand, travelling microscope, a pointer (a pin bent at right angles), detergent.

📐 Result 20.6 — Surface tension from capillary rise

From Jurin's law (Section 7.4, Law 7.21) with angle of contact $\theta\approx0^\circ$ for clean glass and water,

$$S=\frac{\rho g\,r\,h}{2}.$$

A more careful value adds the water in the curved meniscus by replacing $h$ with $h+\tfrac{r}{3}$; for $r$ of a fraction of a millimetre this changes $S$ by less than $1\%$.

Method.

  1. Clean the capillary (with dilute acid or chromic acid, then distilled water) — grease raises the contact angle and spoils the result. Clamp it vertically in the beaker.
  2. Focus the microscope on the bottom of the meniscus in the tube (the cross-wire tangent to it) and read the scale: $R_1$.
  3. Set the pointer so that its tip just touches the water surface outside. Remove the beaker without disturbing the pointer and focus on the pointer tip: $R_2$. Then $h=R_1-R_2$.
  4. Cut the tube at the meniscus level, mount it horizontally and read the two ends of a diameter of the bore with the microscope, in two perpendicular directions; half the mean diameter is $r$.
  5. Add a few drops of detergent to the water, rinse the tube with the solution and repeat. The rise is much smaller.
Liquid$R_1$ (cm)$R_2$ (cm)$h$ (cm)$r$ (mm)$S$ (N/m)
distilled water$9.243$$3.643$$5.600$$0.26$$0.071$
water + detergent$6.243$$3.643$$2.600$$0.26$$0.033$

Precautions and sources of error. The tube must be scrupulously clean and wetted inside (raise and lower the water first, so the meniscus settles on a wet wall); the tube must be vertical; measure $h$ from the bottom of the meniscus; measure the bore exactly where the meniscus stood, since tubes are seldom uniform; note the temperature, because $S$ falls as it rises. The effect of the detergent is a large fall in $S$ (and in $h$), which is why detergents let water wet fabrics.

Experiment 7 — Coefficient of viscosity by terminal velocity

Aim. To determine the coefficient of viscosity of a given viscous liquid (glycerine or castor oil) by measuring the terminal velocity of a falling sphere. Apparatus. Tall, wide glass jar of the liquid, steel ball bearings of different radii, screw gauge, stopwatch, metre scale, thermometer, rubber bands to mark levels.

📐 Result 20.7 — Viscosity from terminal velocity

From Stokes' law and the terminal velocity of a sphere (Section 7.3, Result 7.17), a ball of radius $r$ and density $\rho$ falling at terminal velocity $v_t$ through a liquid of density $\sigma$ gives

$$\eta=\frac{2r^2(\rho-\sigma)g}{9v_t},\qquad v_t\propto r^2 .$$

A graph of $v_t$ against $r^2$ for balls of the same material is a straight line through the origin of slope $\dfrac{2(\rho-\sigma)g}{9\eta}$.

Method.

  1. Measure the diameter of each ball with a screw gauge. Put two rubber bands round the jar, $A$ well below the surface (so that the ball has reached terminal velocity) and $B$ some $20\ \text{cm}$ lower but well above the bottom.
  2. Wet a ball with the liquid and drop it gently along the axis of the jar. Start the stopwatch as it passes $A$ and stop it at $B$; $v_t=AB/t$.
  3. Repeat with balls of several radii. Plot $v_t$ against $r^2$ and use the slope.
  4. Note the temperature of the liquid.
Radius $r$ (mm)$r^2$ (mm²)time $A\to B$ (s)$v_t$ (cm/s)$v_t/r^2$ (cm s⁻¹ mm⁻²)
$1.00$$1.00$$20.0$$1.00$$1.00$
$1.50$$2.25$$8.9$$2.25$$1.00$

Precautions and sources of error. The upper mark must be far enough below the surface for the ball to reach terminal velocity; the jar must be wide (the walls slow the ball) and deep (the bottom too); drop the ball along the axis, wetted, with no air bubble clinging to it; keep the temperature steady — the viscosity of glycerine falls by several per cent per degree. Air bubbles and parallax in timing the marks are the main random errors; the radius enters squared.

✏️ Example 6 — Two balls in glycerine
Steel balls (density $7800\ \text{kg/m}^3$) are dropped in glycerine (density $1260\ \text{kg/m}^3$). A ball of radius $1.00\ \text{mm}$ takes $20.0\ \text{s}$ to fall between two marks $20.0\ \text{cm}$ apart, well below the surface. (a) Find the coefficient of viscosity. (b) How long would a ball of radius $1.50\ \text{mm}$ take between the same marks? (c) A student drew the upper mark just $1\ \text{cm}$ below the surface. Would her value of $\eta$ be too large or too small?
Find $\eta$ in (a) before you open the solution:
(a) $v_t=\dfrac{0.200}{20.0}=1.00\times10^{-2}\ \text{m/s}$. $\eta=\dfrac{2\times(1.00\times10^{-3})^2\times6540\times9.8}{9\times1.00\times10^{-2}}=1.42\ \text{Pa s}$.
(b) $v_t\propto r^2$: $v_t'=1.00\times(1.5)^2=2.25\ \text{cm/s}$, so $t'=\dfrac{20.0}{2.25}=8.9\ \text{s}$.
(c) Near the surface the ball is still speeding up, so its average speed between the marks is less than $v_t$. A smaller $v$ in $\eta=2r^2(\rho-\sigma)g/9v$ gives too large a viscosity.
⏱ Shortcut Same material, same liquid: $t\propto\dfrac{1}{r^2}$. Doubling the radius cuts the time to a quarter.

In the capillary-rise experiment a few drops of detergent are added to the water. Compared with pure water, the rise in the same clean tube

The resonance tube and the method of mixtures

The last two experiments return to waves and heat. The resonance tube uses the standing waves of a closed air column (Section 10.4, Result 10.25); the method of mixtures is the principle of calorimetry (Section 7.5, Law 7.25) — heat lost by the hot body equals heat gained by the cold ones.

Experiment 8 — Speed of sound in air with a resonance tube

Aim. To determine the speed of sound in air at room temperature using a resonance tube and a tuning fork. Apparatus. Resonance tube (a metal or glass tube about $1\ \text{m}$ long, with a reservoir to raise and lower the water level), tuning fork of known frequency (e.g. $512\ \text{Hz}$), rubber pad, metre scale, thermometer.

📐 Result 20.8 — Resonance tube

The air column above the water is a pipe closed at the bottom. With end correction $e$ at the open end, resonance occurs when the column length $l$ satisfies $l+e=\tfrac{\lambda}{4},\tfrac{3\lambda}{4},\dots$ For the first two resonance lengths $l_1$ and $l_2$:

$$\lambda=2(l_2-l_1),\qquad v=2f\,(l_2-l_1),\qquad e=\frac{l_2-3l_1}{2}.$$

The difference $l_2-l_1$ is free of the end correction. The end correction is roughly $0.3\,d$ for a tube of internal diameter $d$.

Method.

  1. Raise the water level nearly to the top. Strike the fork on the rubber pad (never on a hard surface) and hold it horizontally, with a prong vertical, just above the open end.
  2. Lower the water level slowly until the sound swells to a maximum: the first resonance. Fine-tune by moving the level up and down around it; record the column length $l_1$ (from the top to the water surface).
  3. Lower the water further to find the second resonance $l_2$ (about $3l_1$). Repeat each reading three times, approaching from above and below.
  4. Record the room temperature.
$f$ (Hz)$l_1$ (cm)$l_2$ (cm)$\lambda=2(l_2-l_1)$ (cm)$v$ (m/s)$e$ (cm)
$512$$16.0$$49.5$$67.0$$343$$0.75$
$480$$17.2$$53.0$$71.6$$344$$0.70$

Precautions and sources of error. Hold the fork just above (not inside) the tube; locate each resonance by the loudest sound, approaching it from both sides; the tube must be vertical and of uniform bore; strike the fork gently (a hard blow excites overtones). Since $v\propto\sqrt{T}$, the speed found is for the room temperature, which must be recorded; humidity raises it slightly.

Experiment 9 — Specific heat capacity of a solid and of a liquid by the method of mixtures

Aim. To determine the specific heat capacity of a given solid, and of a given liquid, by the method of mixtures. Apparatus. Copper calorimeter with stirrer and lid inside an insulating jacket, two thermometers ($0.1$–$0.5\ ^\circ\text{C}$ least count), steam heater (hypsometer) or boiling-water bath, the solid (metal pieces or a block), the liquid, balance.

📐 Result 20.9 — Method of mixtures

A hot body of mass $m_1$ and specific heat capacity $c_1$ at temperature $T_1$ is put into a calorimeter of mass $m_c$ ($c_c$) holding a liquid of mass $m_2$ ($c_2$) at $T_2$. If no heat is exchanged with the surroundings, at the final temperature $T$:

$$m_1c_1\,(T_1-T)=\left(m_2c_2+m_cc_c\right)(T-T_2).$$

Solid: use water as the liquid ($c_2$ known) and solve for $c_1$. Liquid: use a solid of known $c_1$ and solve for $c_2$.

Method.

  1. Weigh the empty calorimeter with stirrer, then with the cold liquid (about two-thirds full), and note its temperature $T_2$.
  2. Heat the solid in the steam heater until its temperature is steady ($T_1$, near $100\ ^\circ\text{C}$).
  3. Transfer it quickly into the calorimeter, replace the lid, stir gently and record the highest steady temperature $T$.
  4. Weigh again to find the mass of the solid (if not already known) and calculate.
QuantityvalueQuantityvalue
mass of calorimeter (copper, $400\ \text{J kg}^{-1}\text{K}^{-1}$)$0.10\ \text{kg}$initial temperature of water $T_2$$20.0\ ^\circ\text{C}$
mass of water$0.15\ \text{kg}$temperature of hot solid $T_1$$100.0\ ^\circ\text{C}$
mass of solid$0.20\ \text{kg}$final temperature $T$$28.0\ ^\circ\text{C}$

These readings give $c_1=\dfrac{(0.15\times4200+0.10\times400)\times8.0}{0.20\times72.0}=372\ \text{J kg}^{-1}\text{K}^{-1}$ (a brass-like value).

Precautions and sources of error. The main error is heat lost to the surroundings — during the transfer of the hot solid (which cools on the way) and from the calorimeter before the final reading. Transfer quickly, without splashing; use a lagged calorimeter with a lid; stir throughout; start with the water a few degrees below room temperature so that it gains as much heat from the room as it later loses. Heat loss makes $T$ too low, so the calculated specific heat capacity of the solid comes out too small.

🧭 Problem-solving strategy — Questions on experiments
  1. Readings first. Find the least count, then the zero error with its sign (Rules 20.1–20.2), then corrected $=$ observed $-e_0$.
  2. Name the theory. Every experiment is one formula from an earlier chapter; write it down and identify which quantities are measured.
  3. Linearise. For a graph question, rearrange into $y=mx+c$: $l$ against $M$ (Searle), $v_t$ against $r^2$ (viscosity), $A^2$ against $t$ (energy, with $\ln A^2$ linear). The slope or intercept is the answer.
  4. Errors. Add fractional errors, doubling for squared quantities; the smallest measured quantity with the highest power usually dominates.
  5. Systematic effects. Ask which way an uncontrolled effect pushes the result (heat loss, ball not yet at terminal velocity, end correction, zero error).
⚠️ JEE Main traps

1. Negative zero errors. Vernier zero to the left with the $k$-th division coinciding gives $e_0=-(n-k)\,\text{LC}$, not $-k\,\text{LC}$; a screw-gauge zero above the line with the $k$-th division on it gives $-(N-k)\,\text{LC}$. And corrected $=$ observed $-e_0$: a negative error is added.

2. Amplitude versus energy. The pendulum's energy goes as $A^2$: an amplitude down to $50\%$ means energy down to $25\%$. The graph asked for is $A^2$ against $t$.

3. Squared radii. In Young's modulus, capillary rise and viscosity, the radius is small and (in two of them) squared: double its fractional error, and use the radius, not the diameter, in the formula.

4. Resonance tube. $\lambda=2(l_2-l_1)$, not $2l_1$ or $4l_2$; $4l_1$ alone gives $\lambda$ only after adding $4e$. The second resonance is near $3l_1$, not $2l_1$ (a closed pipe has only odd harmonics).

5. Metre-scale balance. If the knife edge is not at the centre of gravity, the scale's own weight has a moment — leaving it out is the commonest wrong answer.

In a resonance-tube experiment with a $480\ \text{Hz}$ fork, the first two resonance lengths are $17.0\ \text{cm}$ and $53.0\ \text{cm}$. The speed of sound and the end correction are

Exercises

Take $g=9.8\ \text{m/s}^2$, the density of water $1.0\times10^3\ \text{kg/m}^3$ and its specific heat capacity $4200\ \text{J kg}^{-1}\text{K}^{-1}$ unless told otherwise. "Coincides" means the Vernier division lines up with a main-scale mark.

I
Vernier callipers and the screw gauge.
  • 1 On a Vernier calliper $20$ VSD $=19$ MSD of $1\ \text{mm}$. With the jaws closed the Vernier zero is to the right of the main zero and the $3$rd division coincides. Find the zero error.
  • 2 The callipers of part 1 are used to find the depth of a test tube: MSR $86\ \text{mm}$, $11$th division coincides. Find the corrected depth.
  • 3 On callipers with $10$ VSD $=9$ MSD of $1\ \text{mm}$, the closed jaws show the Vernier zero to the left of the main zero with the $8$th division coinciding. A beaker gives: outside jaws MSR $68\ \text{mm}$, $5$th division; inside jaws MSR $64\ \text{mm}$, $9$th division. Find its wall thickness.
  • 4 The beaker of part 3 has a corrected internal depth of $80.0\ \text{mm}$. Find its capacity in $\text{cm}^3$.
  • 5 A screw gauge (pitch $1\ \text{mm}$, $100$ divisions) has the zero of its circular scale below the reference line, with the $4$th division on the line, when the faces touch. A wire reads $2\ \text{mm}$ and $63$ divisions. Find its diameter.
  • 6 A screw gauge (pitch $0.5\ \text{mm}$, $50$ divisions) shows the $47$th division on the reference line when closed, with the zero above the line. A sheet reads $1.5\ \text{mm}$ and $12$ divisions. Find its thickness.
  • 7 Five corrected readings of a wire's diameter are $0.84$, $0.86$, $0.85$, $0.83$ and $0.87\ \text{mm}$. Find the mean diameter, the mean absolute error and the percentage error in the area of cross-section.
  • 8 What is backlash error? How is it avoided, and what is the ratchet for?
  • 9 Why are the readings of a wire's diameter taken at several places and in two perpendicular directions at each place?
1 $\text{LC}=1/20=0.05\ \text{mm}$; zero to the right: $e_0=+3\times0.05=+0.15\ \text{mm}$.
2 Observed $86+11\times0.05=86.55\ \text{mm}$; corrected $86.55-0.15=86.40\ \text{mm}$.
3 $\text{LC}=0.1\ \text{mm}$, $e_0=-(10-8)\times0.1=-0.2\ \text{mm}$. $D_\text{ext}=68.5+0.2=68.7\ \text{mm}$, $D_\text{int}=64.9+0.2=65.1\ \text{mm}$. Wall thickness $\tfrac12(68.7-65.1)=1.8\ \text{mm}$.
4 $V=\tfrac{\pi}{4}(6.51)^2(8.00)=266\ \text{cm}^3$.
5 $\text{LC}=0.01\ \text{mm}$, $e_0=+0.04\ \text{mm}$; observed $2.63$, corrected $2.59\ \text{mm}$.
6 $\text{LC}=0.01\ \text{mm}$, $e_0=-(50-47)\times0.01=-0.03\ \text{mm}$; observed $1.62$, corrected $1.65\ \text{mm}$.
7 Mean $0.850\ \text{mm}$; deviations $0.01,0.01,0,0.02,0.02$, mean absolute error $0.012\ \text{mm}$ ($1.4\%$). Area $\propto d^2$: $2\times1.4\%=2.8\%$.
8 Backlash: the lag of a worn screw when its direction of rotation is reversed, so the scale moves before the spindle does. Turn the screw in one direction only while taking a reading. The ratchet slips once the faces grip the object, so every reading is taken with the same gentle pressure and the object is not squeezed.
9 A real wire is neither perfectly uniform along its length nor perfectly circular; averaging over positions and directions gives the mean diameter and reduces random error.
II
Pendulum, metre scale and Young's modulus.
  • 1 A pendulum's amplitude falls from $8.0\ \text{cm}$ to $6.0\ \text{cm}$ in $4$ minutes. What percentage of its energy remains?
  • 2 The pendulum of part 1 has a bob of mass $0.10\ \text{kg}$ and length $1.0\ \text{m}$. Find the energy lost in those $4$ minutes.
  • 3 For another pendulum $A^2$ halves every $3.0$ minutes, starting from $100\ \text{cm}^2$. Find the amplitude after $9.0$ minutes.
  • 4 Why is $A^2$, and not $A$, plotted against time in the pendulum experiment? What does the shape of the graph tell you?
  • 5 A metre scale balances by itself at $50.0\ \text{cm}$. With the knife edge there, a body at the $20.0\ \text{cm}$ mark is balanced by $60\ \text{g}$ at the $85.0\ \text{cm}$ mark. Find the body's mass.
  • 6 A uniform metre scale of mass $120\ \text{g}$ rests on a knife edge at the $30.0\ \text{cm}$ mark. What mass hung at the $5.0\ \text{cm}$ mark keeps it horizontal?
  • 7 In Searle's experiment a wire $1.50\ \text{m}$ long and $0.60\ \text{mm}$ in diameter extends by $0.52\ \text{mm}$ under a load of $2.0\ \text{kg}$. Find Young's modulus.
  • 8 In part 7 the diameter and the extension are each read to $0.01\ \text{mm}$; $L$ and the load are exact. Find the maximum percentage error in $Y$.
1 $(6.0/8.0)^2=0.5625$: $56\%$ remains.
2 $E_0=\tfrac12\dfrac{mg}{L}A_0^2=\tfrac12\times0.98\times(0.080)^2=3.14\times10^{-3}\ \text{J}$; lost $(1-0.5625)E_0=1.37\times10^{-3}\ \text{J}$.
3 Three halvings: $A^2=100/8=12.5\ \text{cm}^2$, $A=3.54\ \text{cm}$.
4 The energy is proportional to $A^2$, so the $A^2$–$t$ graph shows directly how the energy falls. It is a curve, steepest at the start: the energy is lost fastest when the pendulum moves fastest, and equal times remove equal fractions of the energy.
5 $M\times30.0=60\times35.0$, $M=70\ \text{g}$.
6 The scale's weight acts at $50.0\ \text{cm}$, $20.0\ \text{cm}$ from the knife edge: $m\times25.0=120\times20.0$, $m=96\ \text{g}$.
7 $\pi r^2=\pi(0.30\times10^{-3})^2=2.83\times10^{-7}\ \text{m}^2$. $Y=\dfrac{2.0\times9.8\times1.50}{2.83\times10^{-7}\times0.52\times10^{-3}}=2.0\times10^{11}\ \text{Pa}$.
8 $2\times\dfrac{0.01}{0.60}+\dfrac{0.01}{0.52}=3.33\%+1.92\%=5.3\%$.
III
Surface tension and viscosity.
  • 1 A travelling microscope reads the two ends of a diameter of a capillary bore as $3.124\ \text{cm}$ and $3.176\ \text{cm}$. Water rises $5.60\ \text{cm}$ in this tube. Find the surface tension of water (angle of contact $0^\circ$).
  • 2 With a detergent added, the rise in the same tube is $2.60\ \text{cm}$. Find the new surface tension.
  • 3 Why must the capillary tube be cleaned before the experiment, and why is $h$ measured to the bottom of the meniscus?
  • 4 A steel ball (density $7800\ \text{kg/m}^3$) of radius $1.5\ \text{mm}$ falls through an oil (density $900\ \text{kg/m}^3$) at a terminal velocity of $4.0\ \text{cm/s}$. Find the viscosity of the oil.
  • 5 Find the terminal velocity, in the same oil, of a ball of the same steel with radius $1.0\ \text{mm}$.
  • 6 For steel balls in glycerine ($\rho-\sigma=6540\ \text{kg/m}^3$) the graph of $v_t$ against $r^2$ has slope $1.0\times10^4\ \text{m}^{-1}\text{s}^{-1}$. Find the viscosity.
  • 7 Why must the upper mark be well below the surface of the liquid, and why should the jar be wide?
  • 8 On a warm day the balls fall noticeably faster through the same glycerine. Explain, and say why the temperature must be recorded.
1 $d=0.052\ \text{cm}$, $r=0.26\ \text{mm}$. $S=\dfrac{1000\times9.8\times2.6\times10^{-4}\times0.0560}{2}=0.0713\ \text{N/m}$.
2 $S\propto h$: $0.0713\times\dfrac{2.60}{5.60}=0.0331\ \text{N/m}$ — less than half.
3 Grease raises the angle of contact and lowers the rise erratically; Jurin's law with $\cos\theta=1$ needs clean, wetted glass. The column height is defined to the lowest point of the meniscus (the small volume above it is the $r/3$ correction).
4 $\eta=\dfrac{2\times(1.5\times10^{-3})^2\times6900\times9.8}{9\times0.040}=0.845\ \text{Pa s}$.
5 $v\propto r^2$: $4.0\times(1.0/1.5)^2=1.78\ \text{cm/s}$.
6 $\eta=\dfrac{2(\rho-\sigma)g}{9\times\text{slope}}=\dfrac{2\times6540\times9.8}{9\times1.0\times10^4}=1.42\ \text{Pa s}$.
7 The ball must already be at terminal velocity between the marks, or its average speed is too small (and $\eta$ too large). Close walls add drag of their own and slow the ball, so Stokes' law for an unbounded liquid would not apply.
8 The viscosity of a liquid falls steeply as its temperature rises, so $v_t=2r^2(\rho-\sigma)g/9\eta$ is larger. A value of $\eta$ means nothing without the temperature at which it was measured.
IV
Resonance tube and method of mixtures.
  • 1 With a $480\ \text{Hz}$ fork the first two resonance lengths are $17.2\ \text{cm}$ and $53.0\ \text{cm}$. Find the speed of sound.
  • 2 Find the end correction in part 1.
  • 3 A tube of internal diameter $4.0\ \text{cm}$ gives its first resonance at $15.6\ \text{cm}$ with a $512\ \text{Hz}$ fork. Taking the end correction as $0.3d$, find the speed of sound.
  • 4 A resonance tube is $1.00\ \text{m}$ long. With a $512\ \text{Hz}$ fork, $v=340\ \text{m/s}$ and end correction $1.0\ \text{cm}$, how many resonance positions can be found?
  • 5 A brass piece of mass $0.15\ \text{kg}$ at $100\ ^\circ\text{C}$ is dropped into $0.20\ \text{kg}$ of water at $25.0\ ^\circ\text{C}$ in an aluminium calorimeter of mass $0.050\ \text{kg}$ ($c=900\ \text{J kg}^{-1}\text{K}^{-1}$). The final temperature is $29.5\ ^\circ\text{C}$. Find the specific heat capacity of brass.
  • 6 A copper block ($c=400\ \text{J kg}^{-1}\text{K}^{-1}$) of mass $0.20\ \text{kg}$ at $100\ ^\circ\text{C}$ is put into $0.10\ \text{kg}$ of a liquid at $25\ ^\circ\text{C}$ in a copper calorimeter of mass $0.050\ \text{kg}$. The final temperature is $45\ ^\circ\text{C}$. Find the specific heat capacity of the liquid.
  • 7 In part 5, some heat is lost while the brass is carried from the heater to the calorimeter. Is the true specific heat capacity larger or smaller than the value found? Give two precautions that reduce this error.
  • 8 The speed found in part 1 was measured at $25\ ^\circ\text{C}$. Estimate the speed of sound at $0\ ^\circ\text{C}$.
1 $v=2f(l_2-l_1)=2\times480\times0.358=344\ \text{m/s}$.
2 $e=\dfrac{l_2-3l_1}{2}=\dfrac{53.0-51.6}{2}=0.70\ \text{cm}$.
3 $e=1.2\ \text{cm}$; $\lambda=4(l_1+e)=4\times16.8=67.2\ \text{cm}$; $v=512\times0.672=344\ \text{m/s}$.
4 $\lambda=340/512=66.4\ \text{cm}$; $l=\tfrac{\lambda}{4}-e,\ \tfrac{3\lambda}{4}-e,\ \tfrac{5\lambda}{4}-e=15.6,\ 48.8,\ 82.0\ \text{cm}$ (the next, $115\ \text{cm}$, is too long): $3$ positions.
5 Heat gained $=(0.20\times4200+0.050\times900)\times4.5=885\times4.5=3983\ \text{J}$; $c=\dfrac{3983}{0.15\times70.5}=377\ \text{J kg}^{-1}\text{K}^{-1}$.
6 Heat lost by block $0.20\times400\times55=4400\ \text{J}$; gained by calorimeter $0.050\times400\times20=400\ \text{J}$; by liquid $4000\ \text{J}=0.10\,c\times20$, so $c=2000\ \text{J kg}^{-1}\text{K}^{-1}$.
7 The brass enters cooler than $100\ ^\circ\text{C}$, so less heat reaches the water and $T$ is lower; the calculation then gives a $c$ that is too small — the true value is larger. Transfer quickly with a short path; use a lagged calorimeter with a lid (and stir, reading the maximum temperature).
8 $v\propto\sqrt{T}$: $v_0=344\sqrt{\dfrac{273}{298}}=329\ \text{m/s}$.
🎯 Exam corner JEE Main · MCQJEE Main · Numerical

E1. JEE Main · MCQ In a Vernier calliper $10$ Vernier divisions coincide with $9$ main-scale divisions of $1\ \text{mm}$. With the jaws closed, the Vernier zero lies to the left of the main-scale zero and the $7$th Vernier division coincides with a main-scale mark. Measuring the internal diameter of a cup, the main scale reads $4.2\ \text{cm}$ and the $3$rd division coincides. The internal diameter is (A) $4.20\ \text{cm}$ (B) $4.26\ \text{cm}$ (C) $4.30\ \text{cm}$ (D) $4.16\ \text{cm}$

Zero $\text{LC}=0.01\ \text{cm}$; Vernier zero to the left: $e_0=-(10-7)\times0.01=-0.03\ \text{cm}$.
Reading Observed $4.2+3\times0.01=4.23\ \text{cm}$; corrected $4.23+0.03=4.26\ \text{cm}$.
Answer (B).
⏱ Shortcut Left of zero $\Rightarrow$ reads too little $\Rightarrow$ the answer is larger than $4.23$: only (B) and (C) survive, and the error is $3$ (not $7$) least counts.

E2. JEE Main · Numerical In Searle's experiment, a wire of length $2.0\ \text{m}$ and radius $0.50\ \text{mm}$ extends by $0.40\ \text{mm}$ when the load is increased by $3.0\ \text{kg}$. Young's modulus of the material is $x\times10^{10}\ \text{Pa}$. Find $x$, taking $g=10\ \text{m/s}^2$ and $\pi=3.14$. Give the answer to the nearest integer.

Area $\pi r^2=3.14\times(5.0\times10^{-4})^2=7.85\times10^{-7}\ \text{m}^2$.
Modulus $Y=\dfrac{MgL}{\pi r^2l}=\dfrac{3.0\times10\times2.0}{7.85\times10^{-7}\times4.0\times10^{-4}}=1.91\times10^{11}\ \text{Pa}=19.1\times10^{10}\ \text{Pa}$.
Answer $19$.
⏱ Shortcut Only the change in load and the change in extension matter (that is what the slope of the graph uses) — the dead load never enters.

E3. JEE Main · MCQ Pure water rises $6.0\ \text{cm}$ in a clean capillary tube. A detergent lowers the surface tension by $60\%$ (the water still wets the glass), and the solution is placed in a tube of the same glass with half the radius. The new rise is (A) $2.4\ \text{cm}$ (B) $7.2\ \text{cm}$ (C) $9.6\ \text{cm}$ (D) $4.8\ \text{cm}$

Idea $h=\dfrac{2S\cos\theta}{\rho gr}\propto\dfrac{S}{r}$ (density essentially unchanged, $\theta\approx0$).
Ratio $S'=0.40S$, $r'=r/2$: $h'=6.0\times0.40\times2=4.8\ \text{cm}$.
Answer (D).
⏱ Shortcut "Lowers by $60\%$" means multiply by $0.4$, not $0.6$ — (B) is that slip; (A) forgets the radius.

E4. JEE Main · MCQ In a viscosity experiment, a steel ball of radius $1.0\ \text{mm}$ takes $12.0\ \text{s}$ to fall between two marks in glycerine, at its terminal velocity. A ball of the same steel with radius $2.0\ \text{mm}$ takes, between the same marks, (A) $3.0\ \text{s}$ (B) $6.0\ \text{s}$ (C) $1.5\ \text{s}$ (D) $24\ \text{s}$

Idea $v_t=\dfrac{2r^2(\rho-\sigma)g}{9\eta}\propto r^2$, and $t=\dfrac{\text{distance}}{v_t}\propto\dfrac1{r^2}$.
Ratio Doubling $r$ multiplies $v_t$ by $4$: $t=12.0/4=3.0\ \text{s}$.
Answer (A).
⏱ Shortcut (B) assumes $v_t\propto r$; (D) gets the direction wrong — a bigger ball falls faster.

E5. JEE Main · Numerical In a resonance-tube experiment with a tuning fork of frequency $512\ \text{Hz}$, the first resonance is at a column length of $16.0\ \text{cm}$ and the second at $49.5\ \text{cm}$. Find the speed of sound in air in m/s. Give the answer to the nearest integer.

Wavelength $\dfrac{\lambda}{2}=l_2-l_1=33.5\ \text{cm}$, so $\lambda=67.0\ \text{cm}$.
Speed $v=f\lambda=512\times0.670=343.04\ \text{m/s}$. (End correction $e=(49.5-48.0)/2=0.75\ \text{cm}$ — not needed.)
Answer $343$.
⏱ Shortcut Using $4fl_1=328\ \text{m/s}$ is the trap: the single length carries the end correction.

E6. JEE Main · Numerical A metal piece of mass $0.20\ \text{kg}$ at $100\ ^\circ\text{C}$ is dropped into $0.15\ \text{kg}$ of water at $20\ ^\circ\text{C}$ contained in a copper calorimeter of mass $0.10\ \text{kg}$. The final temperature is $28\ ^\circ\text{C}$. Taking the specific heat capacities of water and copper as $4200$ and $400\ \text{J kg}^{-1}\text{K}^{-1}$ and neglecting heat losses, find the specific heat capacity of the metal in $\text{J kg}^{-1}\text{K}^{-1}$. Give the answer to the nearest integer.

Gained $(0.15\times4200+0.10\times400)\times(28-20)=670\times8=5360\ \text{J}$.
Lost $0.20\times c\times(100-28)=14.4\,c$. Setting equal: $c=\dfrac{5360}{14.4}=372.2$.
Answer $372$.
⏱ Shortcut Lump the calorimeter into the water as a "heat capacity" $670\ \text{J/K}$ first; forgetting it gives $350$.
🔬 Research–Teaching NexusResearch-tutored · optional, about 60 minutes

Research question. Is the speed of sound you measure with a resonance tube really the same for every tuning fork — and can a smartphone replace the tuning fork and your ear?

Why it matters. The speed of sound sets the tuning of wind instruments and organ pipes, the design of concert halls, the range finding of ultrasonic parking sensors and sonar, and the calibration of medical ultrasound scanners. Acoustic engineers routinely correct it for temperature and humidity. The resonance tube is the classic way to measure it, and the end correction you meet in it is a real design parameter for flutes and organ pipes.

What to do.

  1. Use three or four tuning forks (for example $256$, $320$, $384$ and $512\ \text{Hz}$), or a free tone-generator app on a phone held just above the tube.
  2. For each frequency find $l_1$ and $l_2$ three times, approaching each resonance from above and from below. If you have a second phone, use a sound-level-meter app to locate the loudest point instead of your ear.
  3. Compute $v=2f(l_2-l_1)$ and $e=(l_2-3l_1)/2$ for each frequency. Record the room temperature.
  4. Plot $l_1$ against $1/f$: from $l_1=\dfrac{v}{4f}-e$ the slope is $v/4$ and the intercept is $-e$.
  5. Compare $e$ with $0.3d$ for your tube, and your $v$ with $331\sqrt{T/273}\ \text{m/s}$.

Tools: Resonance tube, tuning forks or a phone tone generator, metre scale, thermometer, a second phone with a sound-level app (optional), spreadsheet.

Hand in: A table of $l_1$, $l_2$, $v$ and $e$ for each frequency, the $l_1$–$1/f$ graph with its slope and intercept, and a paragraph comparing the methods (ear versus meter, two-resonance versus graph).

Going further. Repeat in a cold room and a warm one (or early morning and afternoon) and test $v\propto\sqrt{T}$. How large a temperature change would you need to see the effect above your experimental scatter?

Instructor note: careful groups get $v$ within about $1\%$ of the temperature-corrected value; the graph method usually gives the most reliable $e$ (typically $0.25d$–$0.35d$). Low-frequency forks give long, broad resonances that are hard to locate by ear — a good discussion of why the method works best near $500\ \text{Hz}$. Phone speakers are not point sources, so their end corrections come out a little larger.

🧪 Self-check quiz

0/5 correct

Q1. Vernier callipers have $10$ Vernier divisions and a least count of $0.1\ \text{mm}$. With the jaws closed, the Vernier zero lies to the left of the main-scale zero and the $4$th Vernier division coincides. The zero error is

Q2. To study the dissipation of energy of a simple pendulum, the graph plotted is

Q3. In the terminal-velocity experiment with balls of the same material, which graph is a straight line through the origin?

Q4. In the resonance-tube experiment, with first and second resonance lengths $l_1$ and $l_2$, the wavelength of the sound is

Q5. In the method of mixtures for a solid, the hot solid loses some heat to the air while it is being transferred. The specific heat capacity calculated (ignoring this) is